1.0 INTENDED USE
This reagent is intended for the quantitative
determination of carbon dioxide (CO2) in serum.
2.0 BACKGROUND
2.1 METHOD AND HISTORY
Early methods for the determination of carbon dioxide
were based on either volumetric or manometric determination of the CO2
released from a sample by acid treatment.
These methods used the instruments of Van Slyke (10.1, 10.2) until they
were replaced by the Natelson microgasometer, (10.3) which still uses
manometric determination of total CO2.
Methods have been developed for Auto Analyzers (10.4) but
these suffer from baseline drift (10.5) and require equipment which many
laboratories do not have.
Enzymatic methods for CO2 have been introduced
by
2.2 TEST PRINCIPLE
PEPC
PEP + HCO3-
-------> Oxalocetate + H2PO4-
MDH
Oxaloacetate
+ NADH + H+ -------> Malate + NAD+
Carbon Dioxide (in the form of bicarbonate ions) reacts
with phosphoenol-pyruvate (PEP), in the presence of phosphoenol-pyruvate
carboxylase (PEPC) to form oxaloacetate and phosphate. The oxaloacetate is then converted to malate
by the action of malate dehydrogenase (MDH) and reduced nicotinamide adenine
dinucleotide (NADH). The decrease in
absorbance at 340nm resulting from the oxidation of NADH is proportional to the
amount of CO2 in the sample.
Interference from endogenous pyruvate and LDH is eliminated by the
inclusion of sodium oxamate.
2.3 CLINICAL SIGNIFICANCE (10.5)
The measurement of CO2 is useful in the
assessment of acid-base balance disturbances.
Elevated CO2 is observed in metabolic alkalosis and
compensated respiratory acidosis. Low CO2
is observed in compensated respiratory alkalosis and metabolic acidosis. Differentiation between the metabolic and
respiratory conditions is only possible through additional laboratory
determinations.
3.0 SPECIMEN COLLECTION AND HANDLING
3.1 PATIENT PREPARATION
No special
patient preparation is required.
3.2 SPECIMEN COLLECTION
Fresh, unhemolyzed serum collected under anaerobic
conditions is the recommended specimen.
Heparinized plasma collected under anaerobic conditions is
acceptable. Oxalate, citrate, and EDTA
should not be used as they cause shifts of electrolytes and water between
plasma and cells. Samples which are
lipemic, icteric, or hemolyzed may be used if a sample blank is performed. (See 5.5 PROCEDURE NOTES.)
Use a
standard venipuncture tube to draw patient sample.
The amount of sample required will depend on the analyzer
used. Call Biotron’s technical service
department at 1-800-5958766 for the recommended sample volume for your
analyzer.
Record the patient's name, date and time of sample
collection and preparation.
3.3 SPECIMEN STORAGE
The sample may be stored in ice water under anaerobic
conditions for up to one hour (10.9.) It
is recommended that testing be done as soon as possible after sample collection
and preparation. Otherwise store the
sample properly using the guidelines above.
4.0 MATERIALS (10 X 10 ml)
Reagents necessary for the determination of carbon
dioxide are included in the kit.
4.1 REAGENT
CO2
reagent contains after reconstitution:
PEP 1.8
mM
Magnesium Sulfate 10
mM
NADH 0.40
mM
MDH (porcine) ³ 1200 U/L
PEPC (microbial) ³ 200 U/L
Sodium oxamate 2.5
mM
Buffer (pH 8.0 ±
0.1) 50
mM
Sodium Azide as
preservative 0.1%
4.2 WARNINGS AND PRECAUTIONS
For In Vitro Diagnostic Use. Not for Internal use in Humans or
Animals. In Vitro Diagnostics reagents
may be hazardous. Avoid ingestion and
skin or eye contact.
The reagent contains sodium azide at 0.1%. This may react with copper or lead plumbing
to form explosive metal azides. Upon
disposal, flush with large volume of water to prevent azide build up.
4.3 REAGENT PREPARATION
Reconstitute the reagent with 10ml of CO2
diluent. Replace the rubber stopper and allow 5 minutes for
reconstitution. Swirl gently until the
contents of the vial are completely dissolved.
Record the date and time of reconstitution.
Note: CO2
diluent should be used as provided. Do
not shake or mix the diluent vial.
4.4 REAGENT STORAGE AND STABILITY
Unreconstituted reagent should be stored at 2-8°C and is
stable until the expiration date on the label.
Do not use the reagent if there is evidence that moisture has entered
the vial, such as caking or incomplete dissolution.
Reconstituted reagent is stable for 24 hours at room
temperature and 7 days at 2‑8° C.
Keep tightly capped at all times and avoid excessive shaking. Do not use the reagent if the absorbance of
the reagent is less than 0.700 at 340 nm.
4.5 ADDITIONAL MATERIALS REQUIRED
4.5.1 Spectrophotometer
or colorimeter capable of reading absorbance at 340 nm.
4.5.2 1 cm cuvettes
or flow cell capable of transmitting light at 340 nm.
4.5.3 Test tubes
capable of holding 2 ml; cylinders for preparing the working reagent.
4.5.4 Pipettes
capable of delivering 1 ml and 10 µl.
4.5.5 Distilled or
deionized water or CO2 diluent for reconstituting the reagent.
4.5.6 Constant
temperature source which can be adjusted to 37° C.
4.5.7 CO2
standard or calibrator.
4.5.8 Normal and
abnormal controls for quality control.
5.0 TEST PROCEDURE
The
following is a general procedure for use on a manual instrument.
5.1 PROCEDURE CONDITIONS
Wavelength 340
nm
Temperature 37°
C
Pathlength 1.0
cm
Mode kinetic
Reaction time 5
min
Sample volume 10
µl
Reagent volume 1.0
ml
Total volume 1.010
ml
Sample to reagent
ratio 1/100
5.2 INSTRUMENT
Any instrument capable of reading absorbance accurately
with a sensitivity of 0.001 absorbance at 340 nm may be used. The band width should be 10 nm or less, stray
light 0.5% or less, and the wavelength accuracy within 2 nm.
5.3 CALIBRATION
Use an aqueous CO2 standard or a serum based
calibrator. The assay is calibrated by
referencing the absorbance of the unknown sample to the absorbance of the
standard (or calibrator).
5.4 PROCEDURE
5.4.1 Prepare the
required volume of working reagent. (See
4.3 Reagent Preparation Section.)
5.4.2 Labels tubes
as "blank", "standard" and "patient".
5.4.3 Add 1.0 ml of
working reagent in each tube.
5.4.3 Pipette 10 µl
of standard (or calibrator) and patient into their respective tubes. Mix, and incubate for 5 minutes 37° C.
5.4.4 Zero the
photometer at 340 nm using distilled water.
5.4.5 Incubate for 45 seconds and record the absorbance at 340
nm of the patient (A1), standard (As1) and the reagent blank (Ab1).
5.4.6 After exactly 30 seconds record the absorbance at 340 nm
of the patient (A2), standard (As2) and the reagent blank (Ab2).
5.5 PROCEDURE NOTES
CO2 from air or the breath of the analyst is a
major interference in this assay.
Reagent preparation, specimen collection, and all storage instructions
must be strictly followed to minimize this interference.
If the sample is very lipemic, icteric or hemolyzed, a
sample blank must be prepared by adding 10 µl of sample to 1 ml of saline. The absorbance of this blank is subtracted
from the absorbance of the respective test and the corrected absorbance is then
used in the calculation.
5.5
CALCULATION AND RESULTS
CO2 mmol/L =
(A1 – A2)
- (Ab1 - Ab2)
‑‑‑-----‑‑‑‑‑‑‑---------------
x concentration of standard (or calibrator)
(As1 –
As2) - (Ab1 – Ab2)
Example:
0.120
CO2
= ‑‑‑‑-----‑‑- X 30 mmol/L = 36 mmol/L
0.100
with (A1 –
A2) - (Ab1 - Ab2) = 0.120
(As1 –
As2) - (Ab1 – Ab2) = 0.100
concentration
of standard = 30 mmol/L
6.0 INTERPRETATION OF RESULTS
6.1 EXPECTED VALUES (10.9)
The range
of expected values is:
23-34 mmol/L
These values are suggested guidelines. It is recommended that each laboratory
establish the normal range for the area in which it is located.
6.2 LIMITATIONS OF PROCEDURE
A number of conditions and substances have been reported
to affect serum Carbon Dioxide levels. (10.10, 10.11, 10.12)
6.3
LINEARITY
This test is linear to 40 mmol/L. Samples exceeding the
linearity limit should be diluted 1:1 with saline, reassayed and the result
multiplied by 2.
7.0 QUALITY CONTROL
Standard practice for quality control should be applied
to this system. Commercially available
lyophilized controls can be used to monitor the daily acceptable
variations. Normal and abnormal controls
should be assayed at the beginning of each run of patient samples, whenever a
new reagent or a different lot number is being used, and following any system
maintenance. A satisfactory level of
performance is achieved when the analyte values obtained are within the
"acceptable range" established by the laboratory.
8.0 CALIBRATION PROCEDURES
Use an aqueous CO2 standard or a serum based
calibrator. The assay is calibrated by
referencing the absorbance of the unknown sample to the absorbance of the
standard (or calibrator). Refer to your
instrument manual for more details.
Calibration is required with the use of a new lot of reagent,
any system maintenance or whenever indicated by quality control data.
9.0 PERFORMANCE CHARACTERISTICS
9.1 PRECISION
Within-Run
Mean
(mmol/L) SD (mmol/L) CV (%)
20.7 0.48 2.3
42.7 0.56 1.3
Between-Run
Mean
(mmol/L) SD (mmol/L) CV (%)
20.1 0.54 2.7
39.7 1.56 3.9
9.2 CORRELATION
A correlation study was done comparing this method (y)
and a commercial reagent using the same methodology (x).
Number Regression Equation Correlation
of
Samples y=Biotron, x=Comparative Coefficient
59 y = 1.10x - 2.4 0.987
10.0 REFERENCES
10.1 Van Slyke,
D.D. and Stadie, W.C.J. Biol. Chem. 49:1 (1921).
10.2 Van Slyke,
D.D. and Neil, J.M.J. Biol. Chem 61:523 (1924).
10.3 Natelson, S.,
Microtechniques of Clinical Chemistry, C. Thomas, Springfield, IL p.147 (1961).
10.4 Skeggs, L.T.
Jr., Am. J Clin. Path. 33:181 (1960).
10.5 Tietz, N.W.,
Fundamentals of Clinical Chemistry, W.B. Saunders, Philadelphia, PA, pp.
884-887 (1982).
10.6 Wilson, W.,
et al, Clin. Chem. 19:640 (1973).
10.7 Menson, R.C.,
et al, Clin. Chem. 20:872 (1974).
10.8 Norris, K.A.,
et al, Clin. Chem 21:1093 (1975).
10.9 Henry, R.J.,
Clinical Chemistry: Principles and Technics, Harper & Row, New York, NY.,
p. 784 (1974).
10.10 Young, D.S.,
et al, Clin. Chem. 21:1D (1975).
10.11 Martin, E.W.,
In Hazard of Medication (Alexander, S.F. Farage, D.J., and Hassan, W.E., Jr.
eds), J.B. Lippincott Co., Philadelphia, PA, and Toronto, Canada, p.169 (1971).
10.12 Constantino,
N.V., and Kabat, H.F., Am J. Hosp. Pharm. 30:24 (1973).
Rev 5/00