SUMMARY
AND EXPLANATION
The Rho(D) antigen was first recognized in 19391. Since the initial recognition
of the D antigen over 40 different antigens are now known to be part of
the Rh system. Most Rh blood group antibodies are immune produced in response
to stimulation by pregnancy or transfusion. The D antigen is highly immunogenic
and has been reported to stimulate the production of anti-D in 50-85%
of D negative individuals who are exposed to D positive blood2. Anti-D
is of considerable importance since this antibody can cause severe Rh
Hemolytic Disease of the Newborn (HDN) and hemolytic transfusion reactions.
The D antigen and its weakened form Du are therefore commonly considered
in the routine selection of blood for transfusion (3 4). Optimal detection
of Du cells by anti-D requires the application of an indirect ant globulin
test procedure. The commonly used terms Rh positive and Rh negative refer
specifically to the presence or absence of the D antigen. The frequency
of Rh positive persons in the general Caucasian population is 85%. A number
of theories that explain the genetic inheritance of Rh antigens have been
proposed and three commonly used systems of Rh terminology have been
developed based on different genetic models. More detailed information
on the Rh system its inheritance and nomenclature may be obtained from
the references cited (2 3 4 5).
BIOLOGICAL PRINCIPLE OF THE TEST
The test used with this blood grouping reagent is based on the principle
of hem agglutination. Incubation of test red cells with TECO Anti-D Blood
Grouping Serum will result in a specific antigen-antibody reaction if
the corresponding D antigen is present on the tested red cells. Visible
detection of this reaction is demonstrated by agglutination of the cells.
No agglutination indicates a negative test result and within the limitations
of the test procedure indicates the absence of the corresponding D antigen.
PRODUCT DESCRIPTION
TECO Anti-D Blood Grouping Serum for slide and rapid tube test is prepared
from human monoclonal Anti-D and human serum containing Anti-D obtained
from immunized donors. TECO Anti-D is designed for use in slide or rapid
tube tests and provides a specific qualitative test for the detection
of the corresponding D antigen on human red blood cells. The diluents used for this reagent contains sodium chloride 6.0 to 8.0% bovine serum
albumin and phosphate buffer. Sodium azide at a final concentration
of 0.1% is added as a preservatives.
CAUTION 1: Sodium azide may react
with copper and lead plumbing to form explosive metal azides. On disposal
flush with large volumes of water to prevent azide build up.
Do not use if marked turbidity or other observable indications of product
alteration occur. These signs may indicate bacterial contamination and/or
product deterioration. Do not use beyond expiration date.
CAUTION 2: Source material from which this product was derived has been
tested in accordance with approved procedures and found non- reactive
for HBsAg and antibody to HIV. No known test method can offer complete
assurance that products derived from human blood will not transmit infectious
agents. Therefore this product should be considered potentially hazardous
and handled with appropriate caution.
SPECIMEN COLLECTION AND PREPARATION
No special preparation of the patient/donor is required prior to specimen
collection. Blood samples should be collected by approved medical procedure.
Blood collected with or without anticoagulant is acceptable if testing
is performed without delay. If a delay in tasting is unavoidable red
calls from clotted samples EDTA or heparin zed ant coagulated samples
should be separated from the serum/plasma washed and stored in a red
cell preservative solution (such as TECO PHYSIOSOL) at 2-8°C for
no longer than 35 days.
ACD CPD and CPDA-1 anticoagulant blood samples may be tested up to
their expiration date if storage is maintained at 2-8°C. Prolonged
storage of red cells prior to testing may result in deterioration of
red cell antigens and resultant weaker than expected test reactions.
MATERIALS AND EQUIPMENT NOT SUPPLIED
1.SLIDE TEST: Glass slides; Pasteur pipettes; centrifuge; applicator
sticks; isotonic saline; illuminated Rh view box with surface temperature
of 40-50°C.
2.RAPID TUBE TEST: Glass test tubes (10x75 mm or 12x75 mm); Pasteur pipettes; centrifuge; isotonic saline.
3.Du TEST: Polyspecific Anti-Human Globulin or Anti-IgG; IgG sensitized
reagent control cells (Coombs Control Cells); 37°C incubator; optical
aid.
4.OTHER RECOMMENDED MATERIAL NOT SUPPLIED: Control cells of known Rh
phenotype.
TEST PROCEDURE
A. SLIDE TEST METHOD
1. Prepare a 40-50% suspension of test red cells. Red cell suspensions
may be prepared in autologous or group compatible serum or plasma.
2. Place one drop of TECO Anti-D (for slide and rapid tube test) on
one end of a labelled prewarmed slide on an illuminated Rh viewbox.
3. Add two drops of the prepared 40-50% suspension of test red cells
to each drop of the TECO Anti-D.
4. Using separate clean applicator sticks thoroughly mix each red cell
suspension over an oval area of approximately 20 x 40 mm.
5. Slowly tilt the Rh viewbox back and forth and examine for macroscopic
hemagglutination. Agglutination may be apparent within a few seconds;
however observation should not continue beyond 2 minutes.
B.
TUBE TEST METHOD
1. Prepare a 3-5% suspension of test red cells. Red cell suspensions
may be prepared in autologous or group compatible serum or plasma or
in isotonic saline.
2. Dispense one drop of TECO Anti-D (for slide and rapid tube test)
into an appropriately labelled test tube.
3. Add one drop of the prepared 3-5% suspension of test red cells to
the test tube.
4. Mix the contents of the test tube thoroughly.
5. Centrifuge for:
a) 30 seconds at 3200-3400 rpm (900-1000 rcf)
b) or centrifugation of equivalent force.
NOTE: The centrifugal force applied should be the minimum required to
produce a clear supernatant and a clearly delineated red cell button
that can be easily resuspended. No single centrifugation speed or time
can be recommended for all types of available centrifuges or test applications.
Centrifuges should be calibrated individually to determine the optimal
time and speed required to achieve the desired results.
6. Gently resuspend red cell button and examine macroscopically for
agglutination. Do not examine microscopically.
7. Grade and record results.
C. Du Test Method:
1. Prepare a 3-5% suspension of test red cells. Red cell suspensions
may be prepared in autologous or group compatible serum or plasma or
in isotonic saline.
2. Dispense one drop of TECO Anti-D (for slide and rapid tube test)
into an appropriately labelled test tube.
3. Dispense one drop of isotonic saline or 6.0 to 8.0% bovine albumin
solution into a second appropriately labelled test tube to serve as
a control.
4. Add one drop of the prepared 3-5% suspension of test red cells to
each tube.
5. Mix contents of both tubes thoroughly and incubate at 37°C for
15 minutes.
6. Wash the cells in each tube three or four times with isotonic saline.
(Decant supernatant saline completely following each wash and ensure
thorough resuspension and mixing of red cells with each new addition
of saline for subsequent washes).
7. Completely decant supernatant saline following last wash and blot
tube edge on a clean absorbent tissue to ensure the removal of all residual
saline and a resultant "dry" red cell button.
8. Add two drops of Polyspecific Anti-Human Globulin or Anti-IgG to
the "dry" button of cells in each tube (Refer to appropriate
manufacturer's Directions for Use for Anti-Human Globulin).
9. Mix gently but thoroughly to resuspend red cell buttons.
10. Centrifuge for:
a) 15 seconds at 3200-3400 rpm (900-1000 rcf)
b) or centrifugation of equivalent force.
NOTE: The centrifugal force applied
should be the minimum required to produce a clear supernatant and a
clearly delineated red cell button that can be easily resuspended. No
single centrifugation speed or time can be recommended for all types
of available centrifuges or test applications. Centrifuges should be
calibrated individually to determine the optimal time and speed required
to achieve the desired results.
11. Gently resuspend red cell buttons and examine macroscopically for
agglutination. Do not examine microscopi-cally.
12. Grade and record results.
QUALITY CONTROL
Appropriate control tests are essential for all laboratory test procedures.
1. If a patient control is run simultaneously with the test and shows
agglutination no valid conclusion concerning the test result can be
reached.
2. The application of IgG sensitized reagent control cells (such as
TECO Coombs Control Cells) is considered an essential control procedure
to confirm the validity of weak or negative antiglobulin tests.
3. It is strongly recommended that the reactivity of Blood Grouping
Sera be confirmed each day of use by control tests with known antigen
positive and negative red cells. Positive control cells should be selected
to represent weak expression of the specific antigen and where applicable
appropriate cells may be selected from heterozygous donors whose red
cells express a single dose of the respective antigen.
INTERPRETATION OF TEST RESULTS
Positive
Negative
POSITIVE (slide and rapid tube test): Agglutination of test red cells
with TECO Anti-D indicates the presence of the corresponding D antigen
(within the accepted limitations of the test procedure).
POSITIVE (Du Test): Agglutination of test red cells with TECO Anti-D
by Du test procedure only (indirect antiglobulin test) indicates that
the red cells are of the Du variety.
NEGATIVE: No agglutination of test red cells with TECO Anti-D indicates
the absence of the corresponding Rh antigen within the accepted limitations
of the test procedure.
LIMITATIONS OF THE TEST PROCEDURE
1. Red cells that have a positive direct antiglobulin test (DAT) may
on rare occasion agglutinate "non-specifically" in tests with
Anti-D Blood Grouping Sera (for slide and rapid tube test). Rarely examples
of test cells may spontane-ously agglutinate in low protein media6.
Such false agglutination may not always be detected by control tests
using saline media6.
2. The use of unwashed test red cells suspended in plasma or serum may
promote false positive reactions associated with rouleaux formation
autoantibodies or more rarely antibodies to constituents of TECO Anti-D
diluent. The routine use of well washed red cells for rapid tube tests
may reduce the incidence of such false positive reactions.
3. Red cells that have a positive direct antiglobulin test cannot be
accurately tested for Du.
4. Some red cells may express variant Rh antigens and may therefore
demonstrate weaker than expected reactions with Anti-D.
5. Delays in reading antiglobulin tests overvigorous resuspension of
red cell buttons and other technique variables associated with antiglobulin
test performance may result in weaker than expected or false negative
Du test results.
6. Rare false positive reactions may occur associated with the presence
of unsuspected antibodies to low incidence antigens. Such antibodies
may infrequently and unavoida-bly occur as contaminants of human blood
grouping reagents.
7. False negative or unexpectedly weak reactions may occur with red
cells that have been subjected to prolonged storage and/or inappropriate
storage conditions.
8. Other variables such as improper technique inappropriate centrifugation
or incubation improperly cleaned glassware and/or contaminated materials
may cause false negative or false positive results.
SPECIFIC PERFORMANCE CHARACTERISTIC
TECO Anti-D (for slide and rapid tube test) has been tested and found
to specifically agglutinate human red cells if the corresponding D antigen
is present when used according to the Recommended Directions for Use.
Each lot of TECO Anti-D has been tested to ensure potency and specificity
according to methods recommended by U.S.-F.D.A. standards.
Specificity tests for TECO Anti-D (slide and rapid tube test) are performed
using extensive panels of Rh negative donor cells that include cells
positive for antigens that occur with a frequency of 1% or greater in
the North American population. In addition each lot of TECO Anti-D is
tested and shown to be non-reactive in serological tests with selected
Bg (a+) donor cells when used according to the Recommended Directions
for Use.
The absence of antibodies to the low frequency antigens Mg and Wra is
not always confirmed; however when test cells are available confirmatory
tests are performed. Antibodies to Lec and Led which are most unlikely
to occur as contaminants in blood grouping serum are not necessarily
excluded.
Deviation from the Recommended Directions for Use may result in less
than optimal product performance.
TECO Anti-D Blood Grouping Serum meets U.S.-F.D.A. and Canadian C.G.S.B.
potency requirements.
REFERENCES
1. Levine P Stetson RE. An unusual case of intragroup agglutination.
J Amer Med Assoc 1939; 113:126-127.
2. Mollison PL. Blood transfusion in clinical medicine. 6th edition.
Oxford Blackwell Scientific 1979.
3. Moore BPL. Serological and immunological methods of the Canadian
Red Cross Blood Transfusion Service. 8th edition. Toronto Hunter Rose
1980.
4. Widmann FK. (Ed.) Technical Manual. 9th edition. Washington D.C.
American Association of Blood Banks 1985.
5. Race RR Sanger R. Blood groups in man. 6th edition. Oxford Blackwell
Scientific 1975.
6. Garratty G. et el. Spontaneous agglutination of red cells with a
positive direct ant globulin test in various media. Transfusion 1984;
24:214-217.
|