1.0 INTENDED USE
This reagent is intended for the quantitative determination of Alkaline Phosphatase in serum.
2.0 BACKGROUND
2.1 METHOD AND HISTORY
Alkaline phosphatase in serum is determined by measuring
the rate of hydrolysis of various phosphate esters under specified
conditions. P-Nitrophenyl Phosphage is
one such phosphate ester and was introduced as a substrate by Fujita in 1939.
(10.2). Bessey, Lowry, and Brock
published an endpoint procedure in 1946 (10.3) while Bowers and McComb reported
a kinetic procedure in 1966 (10.4). The
kinetic procedure has undergone several modifications and been recommended for routine analysis , (10.5,10.6) This liquid reagent is based on the
recommended method of the AACC. (10.7)
2.2 TEST PRINCIPLE
Alk. Phos.
p-Npp + H2O
——————> p-Nitrophenol + H3PO4
p-Nitrophenyl phosphate is hydrolyzed to p-nitrohenol and
inorganic phosphate. The rate at which
the p-Npp is hydrolyzed, measured at 405 nm, is directly proportioned to the
alkaline phosphatase activity.
2.3 CLINICAL SIGNIFICANCE
Serum alkaline phosphatase estimations are of interest in
the diagnosis of two groups of conditions:
hepatobiliary disease, and bone disease associated with increased
osteoblastic activity. (10.1)
3.0 SPECIMEN COLLECTION AND HANDLING
3.1 PATIENT PREPARATION
No special
patient preparation is required.
3.2 SPECIMEN COLLECTION.
Fresh, clear, unhemolyzed serum is the preferred
specimen. EDTA, Oxalate and citrate
inhibit the action of alkaline phosphatase.
Therefore these anticoagulants should be avoided.
Use a
standard venipuncture tube to draw patient sample.
The amount of sample required will depend on the analyzer
used. The amount of serum required is in
the range of 5-25 µl. Call Biotron’s
technical service department at 1-800 595 8766 for the recommended sample
volume for your analyzer.
Record the patient's name, date and time of sample
collection and preparation.
3.3 SPECIMEN STORAGE
Serum for alkaline phosphatase assay may be stored at room temperature (18-26° C) for up to 8 hours. Samples are stable for 4-5 days at 2-8° C and for several months at -10° C. However, it has been reported that increased activities are found after storage (10.8.)
It is recommended that testing be done as soon as
possible after sample collection and preparation. If testing cannot occur immediately, store
the sample properly using the guidelines above.
4.0 MATERIALS
(1X120,1X30 ml)
Reagents necessary for the determination of alkaline phosphatase
are included in the
kit.
4.1 REAGENT
Alkaline Phosphatase working reagent contains:
magnesium acetate >
3.0 mM/L
p‑nitrophenyl
phosphate >
11.0 mM/L
Alkaline Phosphatase buffer contains
AMP buffer
4.1.1 Standard/Control/Calibrator > 0.3
mM/L
4.2 WARNINGS AND PRECAUTIONS
For In Vitro Diagnostic Use. Not for Internal use in Humans or Animals. In Vitro Diagnostics reagents may be hazardous. Avoid ingestion and skin or eye contact.
4.3 REAGENT PREPARATION
The working reagent is prepared by mixing 4 parts of R1 to 1 part of R2.
4.4 REAGENT STORAGE AND STABILITY
Store reagent set at 2-8°C. Unopened reagents are stable until the expiration date is stored as directed. The working reagent is stable for 14 days at 2-8°C, and for 7 days at 18-26°C. Protect from direct light and avoid microbial contamination.
4.5 ADDITIONAL MATERIALS REQUIRED
4.5.1 A
spectrophotometer or colorimeter capable of reading absorbance accurately at
405 nm.
4.5.2 1 cm cuvettes
or a flow cell capable of transmitting light at 405 nm.
4.5.3 Test tubes
and pipettes.
4.5.4 Timer with
one minute increments.
4.5.5 Constant
temperature heat source which can be adjusted to 37° C.
4.5.6 Normal and
abnormal control for quality control.
5.0 TEST PROCEDURE
The following is a general procedure for use on a manual instrument. Application procedures for use on an automated analyzers are available. Contact Biotron’s Technical Service Department for specific information.
5.1 PROCEDURE CONDITIONS
Wavelength 405
nm
Temperature 37°
C
Pathlength 1
cm
Mode Kinetic
Lag Time 1
min.
Sample to reagent
ratio
5.2 INSTRUMENT
Any instrument capable of reading absorbance accurately with a sensitivity of 0.001 absorbance at 405 nm may be used. The band width should be 10 nm or less, stray light 0.5% or less, and the wavelength accuracy within 2 nm.
5.3 CALIBRATION
No reagent calibration is necessary as the alkaline phosphatase activity is calculated by use of the molar absorptivity of p-nitrophenyl which is taken as 18.8 at 405nm.
5.4 PROCEDURE
5.4.1 Prepare the
required number of alkaline phosphatase working reagent. (See 4.3 Reagent Preparation section.)
5.4.2 Into separate
test tubes pipette 25 µl of serum to be assayed.
5.4.3 Add 1 ml of
working reagent. Mix and incubate for 1 minute at 37° C.
5.4.4 Record the
absorbance at one minute intervals until the absorbance change is constant.
5.5
CALCULATION AND RESULTS
Alkaline
Phosphatase U/L =
DA/min X assay volume (ml) X 1000
‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑------------
= DA/min X 2187
18.8 X
light path (cm) X sample volume (ml)
DA/min = change in absorbance per minute
assay
volume = total reaction volume expressed in ml
1000 =
converts U/ml to U/L
18.8 =
absorbance coefficient of p‑nitrophenyl at 405 nm
lightpath
= length of the light path expressed in cm (usually 1)
sample
volume = sample volume expressed in ml
2187 =
factor derived from constants in the equation
Example:
Alkaline
Phosphatase U/L =
0.019 X
1.025 X 1000
‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑‑---------‑ = 0.019 X 2187 = 42 U/L
18.8 X 1 X 0.025
0.019 =
change in absorbance per minute
1.025 =
total reaction volume in ml
1000 =
converts U/ml to U/L
18.8 =
absorbance coefficient of p‑nitrophenyl at 405 nm
1 = light
path in cm
0.025 =
sample volume in ml
Note:
To convert to SI units (nkat/L) multiply
U/L by 16.67.
6.0 INTERPRETATION OF RESULTS
6.1 EXPECTED VALUES (10.4)
The range
of expected values is:
33
‑ 120 U/L (37° C)
These values are suggested guidelines. It is recommended that each laboratory
establish the normal range for the area in which it is located.
6.2 MEDICAL ALERT VALUES
Each laboratory should establish low and high values beyond which the patient would require immediate attention by a physician. If a "medical alert value" is reached, always repeat the test to confirm the result and notify a physician if the result is confirmed.
6.3 LIMITATIONS OF PROCEDURE
A number of substances have been reported to cause physiological changes in serum alkaline phosphatase concentrations. (10.5-10.7)
As with any chemical reaction, users should be alert to
the possible effect on results caused by unknown interferences from medications
or endogenous substances. All patient
results should be evaluated in light of the total clinical status of the
patient.
7.0 QUALITY CONTROL
Standard practice for quality control should be applied to this system. Commercially available lyophilized controls can be used to monitor the daily acceptable variations. Normal and abnormal controls should be assayed at the beginning of each run of patient samples, whenever a new reagent or a different lot number is being used, and following any system maintenance.
A satisfactory level of performance is achieved when the
analyte values obtained are within the "acceptable range" established
by the laboratory.
8.0 CALIBRATION PROCEDURES
No reagent calibration is necessary as the alkaline phosphatase activity is calculated by use of the molar absorptivity of p-nitrophenyl which is taken as 18.8 at 405nm.
9.0 PERFORMANCE CHARACTERISTICS
9.1 PRECISION
The estimates of precision shown below were obtained from assays of human control serum.
Within-Run
Mean
(U/L) SD (U/L) CV
(%)
94 ± 1.93 2.1
319 ± 4.28 1.3
Between-Run
Mean
(U/L) SD (U/L) CV
(%)
95 ±
1.26 1.3
315 ± 3.26 1.2
9.2 CORRELATION
A correlation study was done comparing this method and a similar alkaline phosphatase method. The samples range between 35 and 375 U/L.
Number Regression Equation Correlation
of
Samples y=Biotron,
x=Comparative Coefficient
95 y = 1.2 x + 4.5 0.981
9.3 LINEARITY
This procedure is linear through 1000 U/L beyond which the specimen should be diluted with an equal volume of deionized water. Reassay the specimen and multiply the results by 2.
9.4 SENSITIVITY
The average sensitivity for this method is 0.0003 DA/min per unit of concentration (U/L).
10.0 REFERENCES
10.1
Teitz, N.W.,
Fundamentals of Clinical Chemistry, W.B. Saunders Co., p.603 (1882).
10.2
Fujita, H., J.
Biochem, (
10.3
Bessey, O.A., Lowry,
O.H., Brock, M.J., J. Biol. Chem. 164:321 (1964).
10.4
Bowers, G.N., Jr.,
McComb, R.B., Clin. Chem. 12:70 (1966).
10.5
The Committee on
Enzymes of the Scandinavian Society for Clinical Chemistry and Clinical
Physiology, Scand. J. Clin. Lab. Ivest. 32:291 (1974).
10.6
Wilkinson, J.H., et
al, Clin. Chem. 15:487 (1969).
10.7
Tietz, N.W., et al,
Clin. Chem. 29:751 (1983).
10.8
Young, D.S. et al,
Clin. Chem. 21:1D (1975).
10.9
Demetriou, J.A.
Drewes, P.A.., Gin, J.B., Clinical Chemistry: Principles and Technics, 2nd
Ed., Hagerstown (MD), Haarper & row, p.927 (1974).
10.10 Rej., R., Clin. Chem. 23:1903 (1977).
10.11 NCCLS document “Protection of Laboratory Workers from
Infectious Disease Transmitted by Blood Body Fluids, and Tissue:, 2nd
Ed. (1991).