AST
(SGOT) REAGENT SET
For the quantitative determination of Aspartate Aminotransferase (AST)
in serum.
INTRODUCTION
Serum aspartate aminotransferase (AST) also known as serum glutamic
oxalacetic transaminase (SGOT) is a tissue enzyme that catalyzes the
exchange of amino and keto groups between alpha--amino acids and alpha-keto
acids. AST is widely distributed in tissue principally cardiac hepatic
muscle and kidney. Injury to these tissues results in the release of
the AST (SGOT) enzyme to general circulation. Following a myocardial
infarction serum levels of AST (SGOT) are elevated and reach a peak
48 to 60 hours after onset. Hepatobiliary diseases such as cirrhosis
metastatic carcinoma and viral hepatitis also will increase serum AST
levels. (1)
The
first kinetic assay of AST for diagnostic purposes was described by
Karmen et al. in 1955 using a coupled reaction of malate dehydrogenase
(MDH) and NADH2. This assay system was critically evaluated and optimized
in 1960 by Henry et al.3 In 1977 the International Federation of Clinical
Chemistry recommended a reference procedure for the measurement of AST
activity based upon Karmen's procedures.4 The AST reagent applies the
formulation recommended by the IFCC.
PRINCIPLE
The enzymatic reaction sequence employed in the assay of
aspartate aminotransferase is as follows:
AST
L-Aspartate + 2-Oxoglutarate ------------> Oxalacelate + L-Glutamate
MDH
Oxalacetate + NADH + H+ ---------> L- Malate + NAD+ + H*0
AST
catalyzes the transfer of an amino group between L-aspartate and 2-oxoglutarate.
The oxalacetate formed in the first reaction is then reacted with NADH
in the presence of malate dehydrogenase (MDH) to form NAD. AST activity
is determined by measuring the rate of oxidation of NADH at 340nm. Lactate
dehydrogenase is included in the reagent to convert endogenous pyruvate
in the sample to lactate during the lag phase prior to measurement.
REAGENT COMPOSITION
When reconstituted as directed the reagent for AST contains the following:
(Concentrations refer to reconstituted reagent.)2-Oxoglutarate 12 mM
L-Aspartic Acid 200 mM NADH 0.19 mM LDH 800 U/L MDH 600 U/L Buffer 100
mM pH 7.8 + 0.1. Nonreactive preservatives and fillers.
WARNINGS AND PRECAUTIONS
1. For in vitro diagnostic use.
CAUTION: In vitro diagnostic reagents may be hazardous. Handle in accordance
with good laboratory procedures which dictate avoidng ingestion and
eye or skin contact.
2 Serum specimens should be considered infectious and handled appropriately.
3. Use distilled or deionized water where indicated.
STORAGE AND STABILITY
1. Store dry reagent at 2 - 8°C.
2. The reconstituted reagent is stable for eight (8) hours at room temperature
and for twenty one (21) days when refrigerate immediately.
REAGENT DETERIORATION
The reagent should be discarded if:
1. The initial absorbance read against water at 340 nm is below
0.800.
2. The reagent fails to meet stated parameters of performance.
SPECIMEN COLLECTION
This assay is intended for use with serum. Reports indicate that AST
in serum remains stable at 4°C for a minimum of 7 days.
Hemolyzed specimens should not be used as erythrocytes contain Fifteen
times the AST activity in serum. (5)
INTERFERING SUBSTANCES
Pyridoxal phosphate can elevate AST values by activating the apoenzyme
form of the transaminase. Pyridoxal phosphate may be found in diluent
water contaminated with microbial growth. (6) High levels of um pyruvate
may also interfere with assay performance. Young et al. give a list
of drugs and other substances that interfer with the determination of
AST activity. (7) Refer also to N.E. Saris for a list of references.
(8)
MATERIALS REQUIRED BUT NOT PROVIDED
1. Pipetting devices.
2. Test tubes/rack.
3. Timing device.
4. Spectrophotometer capable of reading at 340 nm (UV).
5. Heating block or bath (37°C).
GENERAL INSTRUCTIONS
The reagent for AST is intended for use either as an automated procedure
on chemistry instruments or as a manual procedure on a suitable spectrophotometer.
PROCEDURE (AUTOMATED)
Consult the appropriate instrument application guide available from
us.
PROCEDURE (MANUAL)
1.Reconstitute Reagent according to instructions on vial label.
2.Pipette 1.0 ml of reagent into a 1cm cuvette and allow to equilibrate
to
37°C.
3.Add 0.10 ml (100 ul) of specimen to reagent and mix gently.
4.Maintain the solution at 37°C. After one (l) minute measure the
absorbance at 340 nm.
5.Take two additional absorbance readings at 1 minute intervals.
Calculate the mean absorbance change per minute. (/\ A/min.)
6.Multiply the /\ A/min. by 1768 to calculate IU/L of AST.
ALTERNATE VOLUMES
If the spectrophotometer being used requires a final volume greater
than 1.0 ml for accurate readings follow the "ALTERNATE PROCEDURE".
CALCULATIONS
One International Unit (IU) is defined as the amount of enzyme that
catalyzes the transformation of one micromole of substrate per minute
under specified conditions.
/\
Abs./min. x TV x 1000 = /\Abs./min. * 1.1 * 1000
e x SV x LD 6.22 x 0.1 x 1
= /\Abs./min. * 176
Where: /\ Abs./min. = Average absorbance change per minute
TV = Total reaction volume (ml)
1000 = Conversion of IU/ml to IU/L
e = Millimolar absorptivity of NADH
SV = Sample volume in ml
LP = Light path in cm
Example: If the average absorbance change per minute = 0.15 then 0.15
x 1768 = 265 IU/L
SI
UNIT: To convert to Sl Units (nkat/L) multiply IU/L by 16.67.
NOTE: If any of the test parameters are altered a new factor must be
calculated using the above formula.
ALTERNATE PROCEDURE
1. Reconstitute reagent according to instructions.
2. Pipette 3.0 ml of reagent into a 1 cm cuvette and allow to
equilibrate to 37°C.
3. Add 0.20 rnl (200 µl) of specimen and mix gently.
4. Maintain the solution at 37°C. After one (1) minute measure
the absorbance (A1) at 340 nm
5. After exactly five (5) minutes read and record absorbance (A2)
6. The difference in absorbance between readings (A1- A2)
multiplied by the factor 514 (see" ALTERNATE PROCEDURE CALCULATIONS")
will yield results in IU/L.
7. Sample with values above 500 IU/L should be diluted 1:1
with saline re-assayed and the results multiplied by two (2).
PROCEDURE NOTES
Turbid or high icteric samples may give readings whose initial absorbance
exceeds the capabilities of the spectrophotometer used. Run this kind
of sample using 0.10 ml (100 µl) sample volume to 3.0 ml reagent
and multiply result by two (2).
ALTERNATE PROCEDURE CALCULATION
One International Unit (IU) is defined as the amount of enzyme that
catalyzes the transformation of one micromole of substrate per minute
under specified conditions.
AST (IU/L) = (A1-A2) x TV x 1000 = (A1-A2) x 3.2 x 1000
/\T x e x SV x LD 5 x 6.22 x 0.2 x 1
= (A1 - A2) x 514
Where: (A1-A2) = Absorbance change
TV = Total reaction volume (ml)
1000 = Conversion of IU/ml to IU/L
AT = Time interval between readings
LP = Light path in cm
e = Millirnolar absorptivity of NADH
SV = Sample volume in ml
Example: If A*= 1.45 and A* = 1.28
then (1.45 - 1.28) = 0.17 x 514 = 87 IU/L
SI UNIT: To convert to SI Units (nkat/L) multiply IU/L by 16.67
NOTE: If any of the test parameters are altered a new factor must
be calculated using the above formula.
QUALITY CONTROL
It is recommended that control be included in each set of assays. Commercially
available control material with established AST values may be used for
quality control. The assigned value of the control material must be
confirmed by the chosen application. Failure to obtain the proper range
of values in the assay of control material may indicate either reagent
deterioration instrument malfunction or procedural errors.
EXPECTED VALUES
Up to 28 IU/L (30°C) Up to 40 IU/L (37°C)
It is strongly recommended that each laboratory establish its own normal
range.
PERFORMANCE CHARACTERISTICS
1. Linearity: 500 IU/L
2. Comparison: A comparison study between the present method with available
commercial product using the same method on 22 fresh serum samples from
12 IU/L to 84 IU/L yielded a coefficient of 0.98 and a regression equation
of
Y = 1.04x - 1.25.
3. Sensitivity: Based on an instrument resolution of A= -0.001 this
procedure has a sensitivity of 2 IU/L
4. Precision studies:
Within Run precision: Two commercial serum controls were assayed twenty
times and the following Within Run precision was obtained.
Within Run
Mean (lU/L) S.D. C.V.
22.3 1.1 4.8%
88.3 5.1 5.7%
Run
to Run Run precision:
Two commercial serum controls were assayed for five consecutive days
(triplicate for each level) the following Run to Run precision was obtained.
Run to Run
Mean IU/L S.D. C.V.
21.9 2.3 10.5%
83.8 3.6 4.4%
TEMPERATURE CONVERSION FACTOR (Tf)
(3)
Assay Mixture Tf 25°C Tf 30°C Tf 32°C Tf 37°C
25°C 1.00 1.37 1.57 1.96
30°C 0.78 1.00 1.13 1.43
32°C 0.65 0.89 1.00 1.27
37°C 0.51 0.70 0.73 1.00
Example: If the reaction is performed at 30°C but is to be reported
at 37°C simply multiply the result obtained at 30°C by the factor
1.43 to obtain a correct value.
NOTE: Since temperature factors give only an approximate conversion
it is suggested that values be reported at the temperature of the measurement.
REFERENCES
1. Henry J.B.: Clinical Diagnosis and Management by
Laboratory Methods. W.B. Saunders and Co. Philadelphia
PA. p 361 (1974).
2. Karmen A. et al.: J. Clin. Invest 34:126 (1955).
3. Henry R.J. et al.: Am. J. Clin. Path 34:381 (1960).
4. The Committee on Enzymes of the Scandinavian Society for
Clinical Chemistry and Clinical Physiology. Scand. J. Clin.
Lab.Invest32:291(1974).
5. Henry R.J.: Clinical Chemiatry Principles and Techniques
2nd Ed. Harper and Row New York p. 882 (1974).
6. Reg R. et al.: Clin. Chem. 19:92 (1973).
7. Young D.S.etal.:Clin.Chcm.21:5(1975).
8. Saris N.E. (ed):Clin.Chem23:887(1977).
9. Tietz N.W.: Fundamentals of Clinical Chemistry W.B.
Saunders Co. Phila. p. 682 (1976).
|