INTENDED USE
For the quantitative determination of alanine aminotransferase in serum.


INTRODUCTION
The enzyme alanine aminotransferase is widely reported in a variety of tissue sources. The major source of ALT is of hepatic origin and has led to the application of ALT determinations to the study of hepatic diseases. Elevated serum levels are found in hepatitis cirrhosis and obstructive jaundice. Levels of ALT are only slightly elevated in patients following a myocardial infarction. (l)

UV methods for ALT determination were first developed by Wroblewski and LaDue in 1956. (2) The method was based on the oxidation of NADH by lactate dehydrogenase (LDH). In 1980 the International Federation of Clinical Chemistry recommended a reference procedure for the measurements of ALT based on the Wroblewski and LaDue procedure. (3) The ALT reagent conforms to the formulation recommended by the IFCC.


PRINCIPLE

The enzymatic reaction sequence employed in the assay of ALT is as follows:
ALT
L-Alanine + 2-oxoglutarate --------> Pyruvate + L-Glutamate
LDH
Pyruvate + NADH + H+ ---------> Lactate + NAD+ + H20

The pyruvate formed in the first reaction is reduced to lactate in the presence of lactate dehydrogenase and NADH. The activity of ALT is determined by measuring the rate of oxidation of NADH at 340 nm. Endogenous sample pyruvate is converted to lactate by LDH during the lag phase prior to measurement


REAGENTS
After combining R1 (Buffer Reagent) and R2 (Co-Enzyme) the reagent contains:

L-Alanine 500 mmol/L
LDH >1200 U/L
Tris Buffer pH 7.5 100 mmol/L
2 - Oxoglutarate 15 mmol/L
NADH (Disodium salt) 0.18 mmol/L
Stabilizers and Preservatives


PRECAUTIONS
The reagents are for "In Vitro Diagnostic Use". Normal precautions exercised in handling laboratory reagents should be followed. The reagents contain sodium azide which may be toxic if ingested. Sodium azide may also react with lead and copper plumbing to form highly explosive metal azides. Refer to Material Safety Data Sheet for any updated risk hazard or safety information.


REAGENT PREPARATION
The working reagent is prepared by mixing five (5) volumes of R1 with one (1) volume of R2 in a disposable container.
Example: 25ml R1 + 5ml R2


REAGENT STORAGE AND STABILITY
Reagents are stable until the expiration date on their respective labels when properly stored at 2-8°C and protected from light. Reagents should appear clear and colorless. Discard if either appears cloudy or contains particulate matter. The Working Reagent is stable for 2 weeks at 2-8°C. The Working Reagent should be discarded if the initial absorbance read against distilled water at 340 nm is below 1.000.


MATERIAL REQUIRED BUT NOT PROVIDED
1) Spectrophotometer capable of absorbance reading at 340 nm and
1 cm lightpath
2) Constant temperature block or bath 37°C or temperature controlled
cuvette well
3) Accurate pipetting devices
4) Test tubes
5) Interval timer


SPECIMEN COLLECTION AND STORAGE
Non-hemolyzed serum is the specimen of choice. Whenever possible specimens should be separated and analyzed on the day of collection. Store serum in stoppered tubes. About 10% ALT is lost 3 days at 4°C and in 1 day at 25°C. (5)


INTERFERING SUBSTANCES
Hemolysis must be avoided as the concentration of ALT in red cells is roughly 5 times that of serum. (2) Bilirubin levels up to 40 mg/dL and triglyceride levels up to 2000 mg/dL show no interference in this test. Certain drugs and other substances are also known to affect ALT values. (5)


MANUAL PROCEDURE
1. Prepare ALT Working Reagent according to instructions.
2. Pipette 1.0 mL of working reagent into tubes labeled "controls" patient(s)" etc.
3. Pre-incubate all tubes at 37°C for at least five minutes
4. Zero spectrophotometer at 340 nm with distilled water.
5. Add 100 µL (0.10 mL) serum to its respective tube mix gently and return to a thermo cuvette.
6. Read and record absorbance at 1 minute. Continue incubating at 37°C and record absorbance again at 2 and 3 minutes. Rate should be constant.
7. Determine the average absorbance per minute (DA/min) multiply by factor 1768 for results in U/L.
8. Repeat the procedure for each sample.

NOTE: If cuvette is not temperature controlled incubate samples at 37°C between readings.


AUTOMATED PROCEDURE

Instrument Application Procedure describing the use of this reagent on automated analyzers are available. Please contact Teco's customer service department.


QUALITY CONTROL
It is recommended that controls be included in each set of assays. Commercially available control material with established ALT values may be routinely used for quality control. The assigned value of the control material must be confirmed by the chosen application. Failure to obtain the proper range of values in the assay of control material may indicate either reagent deterioration instrument malfunction or procedure errors.


CALIBRATION
ALT activity is based on the "micromolar extinction coefficient" of NADH at 340 nm (see "Results" section). The instrument manufacturer's calibration guidelines should be followed to calibrate your analyzer. Assaying the ALT contents of a control serum with known ALT values can be used to assure instrument calibration has been performed correctly.


RESULTS
Values are derived based on the "absorptivity micromolar extinction coefficient" of NADH at 340 nm (0.0063). Units per liter (U/L) of ALT/GPT activity is that amount of enzyme which oxidizes one µmol/L of NADH per minute.

/\A/Min Total Volume
U/L = Absorbtivity x Sample Volume /\A/Min 1.100
U/L = 0.0063 x 0.100 .
.
U/L = /\A/Min x 1746


LIMITATIONS
If the DA/min. is greater than 0.342 dilute 1 part sample with 9 parts isotonic saline and re-assay. Multiply the result by 10. ALT values for neonatal patients have not been established with this procedure. Grossly icteric or turbid specimen may require the use of a sample blank.


EXPECTED VALUES

Normal Range: 3 - 35 U/L (37°C)

This range should serve only as a guideline. It is recommended that each laboratory establish its own range of expected values since differences exist between instruments laboratories and local populations.


PERFORMANCE CHARACTERISTICS

Comparison: A group of 128 sera ranging in ALT activity from 7 - 625 U/L was assayed by the described ALT method and by a similar commercially available ALT reagent. Comparison of the results yielded a correlation coefficient of 0.999 and the regression equation was y = 0.960x + 3.2. (Comparison studies were performed according to NCCLS Tentative Guideline EP9-T.)

Precision:
Within-Run
Serum 1 Serum 2
Mean ALT (U/L) 25.1 116.0
Std. Deviation (U/L) 0.82 0.90
C.V. (%) 3.25 0.77

Total Precision
Serum 1 Serum 2
Mean ALT (U/L) 25.8 114.8
Std. Deviation (U/L) 1.13 0.8
C.V. (%) 4.40 0.69

Precision studies were performed according to NCCLS Tentative Guideline EP5-T.

Linearity: Linear to 500 U/L at 37°C. Performed according to NCCLS Guideline EP6-P.

Sensitivity: Based on an instrument resolution of A = 0.001 the method presented shows a sensitivity of 1.8 U/L.


REFERENCES
1. Henry J.B.: Clinical Diagnosis and Management by Laboratory Methods W.B. Saunders and Co. Philadelphia PA p-332-335 (1974)
2. Wroblewski F. and LaDue J.S. Proc. Soc. Exper. Biol. and Med. 91:569 (1956)
3. International Federation of Clinical Chemistry J. Clin.
Chem.Clin.Bio.18:5231(1980).
4. Bergmeyer H.U. Principles of Enzymatic Analysis. Verlag Chemic 1978.
5. Young D.S. Effects of drugs on clinical laboratory tests. AACC Press Washington D.C. 1990
6. Henry JB. Clinical Diagnosis and Management by Laboratory Methods 17th ed. WB Saunders Co. 1984 p 1437.