1.0 INTENDED USE
This reagent is intended for the quantitative
determination of alanine aminotransferase (ALT) in serum.
2.0 BACKGROUND
2.1 METHOD AND HISTORY
Henley and Pollard (10.1) and Wroblewski and Ladue (10.2) developed kinetic procedures for assaying alanine aminotransferase, formerly called glutamate pyruvate transaminase (GPT). The procedures were based on the oxidation of NADH by lactate dehydrogenase (LDH). Henry et. al.(10.3) developed an optimized method and this procedure is a modification of the method of Henry. The NADH concentration and LDH concentration are increased to improve linearity and completely remove interference from endogenous pyruvate in the sample.
2.2 TEST PRINCIPLE
The ALT catalyzes the conversion of L‑alanine and 2‑oxoglutarate
to pyruvate and L‑glutamate. Then LDH catalyzes the oxidation of NADH to
NAD.
ALT
2‑oxoglutarate
+ L‑alanine --------> L‑glutamate + Pyruvate
LDH
Pyruvate +
NADH + H+ ---------> Lactate + NAD+
The rate of decrease in absorbance of the reaction
mixture at 340 nm, due to the oxidation of NADH is directly proportional to the
ALT activity.
2.3 CLINICAL SIGNIFICANCE
The activity of alanine aminotransferase (ALT, GPT) in
serum is clinically useful in the differentiation of hepatic disorders,
especially when its activity is compared to that of aspartate aminotransferase
(AST, GOT.) In cases of hepatic
necrosis, the activity of serum ALT is typically greater than that of AST, whereas
in cirrhosis and metastatic carcinoma, the reverse often occurs. In cases of myocardia infarction, however,
ALT levels may be within the reference interval while AST may frequently be
20-30 times greater.
3.0 SPECIMEN COLLECTION AND HANDLING
3.1 PATIENT PREPARATION
No special
patient preparation is required.
3.2 SPECIMEN COLLECTION
Fresh, clear, unhemolyzed serum is the preferred
specimen. No interference is experienced
with plasma from commonly used anticoagulants.
Use a
standard venipuncture tube to draw patient sample.
The amount of sample required will depend on the analyzer used. The amount of serum required is in the range of 5-200 µl. Call Biotron's technical service department at 1-800-595 8766 for the recommended sample volume for your analyzer.
Record the patient's name, date and time of sample
collection and preparation.
3.3 SPECIMEN STORAGE
Specimens for analysis should be stored at 2 to 8°C
(refrigerated) and are stable for up to 7 days.
Specimens may be stored at -20 to 0°C (frozen) for longer periods.
It is recommended that testing be done as soon as
possible after sample collection and preparation. If testing cannot occur immediately, store
the sample properly using the guidelines above.
4.0 MATERIALS (10 X 10 ml)
(6 X 50 ml)
(6 X 500 ml)
Reagents
necessary for the determination of ALT are included in the kit.
4.1 REAGENT
ALT reagent contains, after reconstitution with deionized
water:
NADH ³ 0.4 mM
lactate
dehydrogenase (lactobacillus leichmannii) ³ 800 U/L
2-oxoglutarate 13
mM
dL-alanine 13
mM
tris buffer 100
mM (pH = 7.3)
preservative
4.2 WARNINGS AND PRECAUTIONS
For In Vitro Diagnostic Use. Not for Internal use in Humans or
Animals. In Vitro Diagnostics reagents
may be hazardous. Avoid ingestion and
skin or eye contact.
4.3 REAGENT PREPARATION
Add 10 ml of the deionized water to each of the required
number of vials of ALT Reagent. Replace
the rubber stopper and allow 5 minutes for reconstitution. Swirl gently until the contents of the vial
are completely dissolved. Record the
date and time of reconstitution.
4.3.2
Add 50 ml of the deionized water to each of the required
number of vials of ALT reagent. Replace
the rubber stopper and allow 5 minutes for reconstitution. Swirl gently until the contents of the vial
are completely dissolved. Record the
date and time of reconstitution.
.
4.3.3
Add 500 ml of the deionized water to each of the required
number of vials of ALT reagent. Replace
the rubber stopper and allow 5 minutes for reconstitution. Swirl gently until the contents of the vial
are completely dissolved. Record the
date and time of reconstitution.
4.4 REAGENT
STORAGE AND STABILITY
Unopened reagents included in the kits are stable at 2-8°
C (refrigerated) until the expiration date stated on the labels. The reconstituted reagent is stable at 2-8° C
(refrigerated) for 14 days or at 18-26° C (room temperature) for 8 hours.
The reagent solutions should be clear. Cloudiness indicates contamination and the
reagent should be discarded. The initial
absorbance of the reagent read against distilled water at 340 nm should be at
least 1.100 to be considered suitable for use.
4.5 ADDITIONAL MATERIALS REQUIRED
4.5.1 Spectrophotometer
capable of reading absorbance at 340 nm.
4.5.2 1 cm cuvettes
or a flow cell capable of transmitting light at 340 nm.
4.5.3 Test tubes
capable of holding 4 ml.
4.5.4 Pipettes
capable of delivering 3.0 ml and 200 µl; cylinders capable of measuring 10 ml.
4.5.5 Timer with
one minute increments.
4.5.6 Constant
temperature source which can be adjusted to 30°C or 37°C.
4.5.7 Normal and
abnormal control for quality control.
5.0 TEST PROCEDURE
The
following is a general procedure for use on a manual instrument.
5.1 PROCEDURE CONDITIONS
Wavelength 340
nm
Temperature 30°C
or 37°C
Pathlength 1.0
cm
Mode Kinetic
Reaction time 2
‑ 4 min
Sample volume 200
µl
Reagent volume 3.0
ml
Total volume 3.2
ml
Sample to reagent
ratio 1/15
5.2 INSTRUMENT
Any instrument capable of reading absorbance accurately
with a sensitivity of 0.001 absorbance at 340 nm may be used. The band width should be 10 nm or less, stray
light 0.5% or less, and the wavelength accuracy within 2 nm.
5.3 CALIBRATION
No reagent calibration is necessary as this procedure is
standardized based on the millimolar absorptivity of NADH which is taken as
6.22 at 340 nm under the test conditions described.
5.4 PROCEDURE
5.4.1 Prepare the
required volume of ALT working reagent.
(See 4.3 Reagent Preparation section.)
5.4.2 Into separate
test tubes pipette 200 µl of serum to be assayed.
5.4.3 Add 3.0 ml of
working reagent, mix, and incubate for one to two minutes at 30°C or 37°C. The lag time will be decreased if the reagent
is prewarmed to the incubation temperature.
5.4.4 Record the
decrease in absorbance at 340 nm at one minute intervals until the absorbance
change is constant.
5.5 CALCULATION AND RESULTS
ALT (U/L)
=
DA/min X assay volume (ml) X 1000
--------------------------------------------------------
= DA/min X 2572
6.22 X
light path (cm) X sample volume (ml)
DA/min = change in absorbance per minute
Assay
volume = total reaction volume expressed in ml
1000 =
converts U/ml to U/L
6.22 =
absorbance coefficient of NADH at 340 nm
Light path
= length of the light path expressed in cm (usually 1)
Sample
volume = sample volume expressed in ml
2572 =
factor derived from constants in the equation
Example: ALT (U/L) =
.017 X 3.2
X 1000
---------------------------
= .017 X 2572 = 44 U/L
6.22 X 1 X 0.2
0.017 =
change in absorbance per minute
3.2 =
total reaction volume expressed in ml
1.0 =
length of the light path expressed in cm
0.2 =
sample volume expressed in ml
6.0 INTERPRETATION OF RESULTS
6.1 EXPECTED VALUES (10.5)
The range
of expected values is:
8‑20 U/L
(30° C)
12‑31 U/L
(37° C)
These values are suggested guidelines. It is recommended that each laboratory
establish the normal range for the area in which it is located.
6.2 MEDICAL ALERT VALUES (10.7)
Each laboratory should establish low and high values
beyond which the patient would require immediate attention by a physician. If a "medical alert value" is
reached, always repeat the test to confirm the result and notify a physician if
the result is confirmed.
6.3 LIMITATIONS OF PROCEDURE
This procedure measures total ALT. Red blood cells contain high concentrations
of ALT, therefore, hemolysis can elevate results. A summary of the influence of drugs on
clinical laboratory test may be found by consulting Young, D.S., et. al. (4).
7.0 QUALITY CONTROL
Standard practice for quality control should be applied
to this system. Commercially available
lyophilized controls can be used to monitor the daily acceptable
variations. Normal and abnormal controls
should be assayed at the beginning of each run of patient samples, whenever a
new reagent or a different lot number is being used, and following any system
maintenance.
A satisfactory level of performance is achieved when the
analyte values obtained are within the "acceptable range" established
by the laboratory.
8.0 CALIBRATION PROCEDURES
No reagent calibration is necessary as this procedure is
standardized based on the millimolar absorptivity of NADH which is taken as
6.22 at 340 nm under the test conditions described.
9.0 PERFORMANCE CHARACTERISTICS
9.1 PRECISION
The estimates of precision shown below were obtained from assays of human control serum.
Within-Run
In this
study, 30 replicates of 2 control sera were run.
Mean
(U/L) SD (U/L) CV (%)
25.0 ±1.4 5.6
84.5 ±4.0 4.6
Between-Run
In this
study, two control sera were assayed twice a day for 10 days.
Mean
(U/L) SD (U/L) CV (%)
44.6 ±2.7 6.0
77.8 ±4.4 5.6
9.2 CORRELATION
A correlation study was done comparing this method (y) a
similar UV alanine aminotransferase procedure (x). 47 samples with a range from 6 to 109 U/L
were assayed. Linear regression analysis
gave the following result.
Number of Regression Equation Correlation
Samples y=Biotron, x=Comparative Coefficient
47 y = 1.008 x - 1.038 .977
9.3 LINEARITY
This procedure is linear to 350 U/L. Procedures on automated instruments which use
greater than one to fifteen dilution factor will have an extended
linearity. A sample with an alanine
aminotransferase activity exceeding the linearity limit should be diluted with
0.9% saline and reassayed incorporating the dilution factor in the calculation
of the result.
9.4 SENSITIVITY
An absorbance change of 0.001 DA/min corresponds to approximately 2.6 U/L ALT activity.
10.0 REFERENCES
10.1 Henley, K.S.,
Pollard, H.M., J.Lab. and Clin.Med.,46,785(1955)
10.2 Wroblewski,
F., Ladue, J.S., Proc. Soc. Exp. Biol. Med. 91,569(1956)
10.3 Henry, R.J.,
Chiamori, N., Gobub, O.J. and Berkman, S.,
Am. J. Clin. Pathology, 34,381(1960)
10.4 Young, D.S.,
Effects of Drugs on Clinical Laboratory Tests, 3rd ed., Washington DC, AACC
Press (1990).
10.5 Tietz, N.W.
(editor) Clinical Guide to Laboratory Tests (1983), W.B. Saunders Company,
Toronto. p. 16.
10.6 Clinical
Diagnosis, 15th ed., J. Davidson and J.B. Henry, Eds, W.B. Saunders Company,
Philadelphia, PA 1974.
10.7 G.J. Kost,
"Critical Limits for Urgent Clinician Notification at U.S. Medical
Centers"; JAMA, Feb. 2, 1990; Vol 263, No.5, p.704